Contact an Account ManagerQuantification of N-(3-chloro-2-hydroxypropyl)valine in human haemoglobin as a biomarker of epichlorohydrin exposure by gas chromatography-tandem mass spectrometry with stabl-isotope dilution
Summary
Type: GC
Application #: 213700
Catalog#: 726205.30 - GC Column, Optima-1 MS, 0.25mm, 0.25um, 30m
Detection:
Conditions: Injection volume: 1 µL, splitless
Injection temperature: PTV injector 120 °C --> 320 °C, 10 °C/min (4 min)
Carrier gas: helium, 1.0 mL/min
Temperature: 80 °C (1 min) --> 340 °C, 8 °C/min (2 min)
Authors: Bader, M. et al.
Source: J. Chromatogr. B 877 (2009), 1402–1415
Details